Strany 1 - Inform II

Inform IIQuick Reference Guidefor Healthcare ProfessionalsBLOOD GLUCOSE MONITORING SYSTEM

Strany 2

10Meter, Front ViewACCU-CHEK Inform II MeterTest Strip PortTouchscreenOn/Off Button

Strany 3 - Table of Contents

11Meter, Back ViewACCU-CHEK Inform II MeterACCU-CHEK Inform II MeterBarcode Scanner WindowBattery PackReset ButtonInfrared Window

Strany 4

12ACCU-CHEK Inform II MeterThe base unit:•Chargesthemeterbatterypack.•Communicateswithadatamanagement system, if available.Base Unit

Strany 5 - Important Safety Information

13Patient TestingNote: Prior to testing, ask patient to wash hands with soap and water, then let dry. If patient is unable to wash hands, wipe with al

Strany 6 - General Information

144. Touch Patient Test to display the Patient ID screen.How to Perform Patient Testing

Strany 7

155. Enter or barcode scan patient ID, then touch . You are prompted to confirm the test strip lot.Patient TestingHow to Perform Patient Testing

Strany 8

166. Verify the test strip lot.•If correct, touch . You are prompted to insert the test strip.•If incorrect, touch to select a different test str

Strany 9

177. Slide the test strip into the test strip port as far as it will go in the direction indicated by the arrows on the test strip.8. Wait until the

Strany 10 - ACCU-CHEK Inform II Meter

189. Apply the drop of blood to the front edge (yellow dosing area) of the test strip. Do not apply the blood to the top of the strip.Blood is pulled

Strany 11

1910. When the test is completed, the result is displayed. Touch to enter the desired comments. Touch to return to the Main Menu.How to Perform Pat

Strany 13 - Patient Testing

2011. Remove the test strip and dispose of it in accordance with applicable regulations and directives for disposal of potentially infectious samples

Strany 14 - Patient ID screen

21Control TestingHow to Perform Control TestingGlucose control tests should be run in the following circumstances:•Before using the meter for the pat

Strany 15 - ID, then touch . You are

22How to Perform Control Testing•If questionable test results are displayed repeatedly•If you wish to test the performance of the systemIn addition,

Strany 16

231. Press the On/Off button to power on the meter.2. Once the Power Up screen appears, touch to proceed to the Operator ID screen, or wait 5 sec

Strany 17

243. Enter or barcode scan operator ID, then touch to display the MainMenu screen.4. Touch Control Test to display the Control Test screen.How to

Strany 18

255. Touch the control level to be run.How to Perform Control TestingControl Testing

Strany 19

266. Verify the control lot number.•If correct, touch . You are prompted to confirm the test strip lot.•If incorrect, touch to enter or barcode

Strany 20

277. Verify the test strip lot.•If correct, touch . You are prompted to insert the test strip.•If incorrect, touch to select a different test st

Strany 21 - Control Testing

288. Slide the test strip into the test strip port as far as it will go in the direction indicated by the arrows on the test strip.9. Wait until the

Strany 22

2910. Apply a drop of glucose control solution to the front edge (yellow dosing area) of the test strip. Do not apply the control solution to the top

Strany 23

3Table of ContentsImportant ... 4General Information ... 6ACCU-CHEK Inform II Met

Strany 24

3011. When the test is completed, the result is displayed. Touch to enter the desired comments. Touch to continue with the next level, if necessary

Strany 25

3112. Remove the test strip and dispose of it in accordance with applicable regulations and directives.How to Perform Control TestingControl Testing

Strany 26

321. Press the On/Off button to power on the meter.2. Once the Power Up screen appears, touch to proceed to the Operator ID screen, or wait 5 sec

Strany 27

33Storing Test Strip Lot Information4. Touch Strip Lots to open the related menu.5. Touch Add if you want to add the information for a new test strip

Strany 28

346. Insert the new code key in the opening of the code key reader. An LED starts flashing green and signals that the reader is ready to transfer dat

Strany 29

357. Hold the meter 4-6 in above the code key reader so that a connection can be made between the two infrared windows.8. Touch to begin downloadin

Strany 30

369. Confirm the test strip lot information.•Touch to store the data for this lot number without changes, or•Touch to modify the data for this te

Strany 31

37The meter does not power on.Solution: Confirm that the meter battery pack is charged correctly.The “QC Due: Immediately” message appears on the scre

Strany 32

05234654001 2012-10 USAACCU-CHEK, ACCU-CHEK INFORM, and COBAS are trademarks of Roche.©2012 Roche Diagnostics. All rights reserved. Roche Diagnostic

Strany 33

4WARNING: Protection against infection• Refer to the operator's manual for the proper cleaning and disinfection instructions for this system.•

Strany 34 - An LED starts flashing green

5WARNING: Protection against infection (continued)• Do not dispose of the battery pack with normal domestic waste. As a component of the ACCU-CHEK In

Strany 35

6General Information• Refer to the ACCU-CHEK Inform II System Operator’s Manual and package inserts for detailed information. Be sure to read the “I

Strany 36

7General InformationYour system is ready to use when it has been charged in the base unit.Meters can be configured according to your institution’s nee

Strany 37 - Troubleshooting

8QC Lockout occurs when •patient testing is attempted and controls have not been run in the time interval or frequency designated by your institutio

Strany 38 - 05234654001 2012-10 USA

9ReagentsYou need the following reagents to perform patient tests and glucose control tests:•ACCU-CHEK Inform II Test Strips•ACCU-CHEK Inform II Con