Roche ACCU-CHEK Inform II Návod k obsluze Strana 147

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 192
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 146
General Product Information
147
Subject/Attribute Range Default Device DMS
Strip lot verication 0: display only
1: yes/no conrmation
2: list selection
3: scan only
1 Y Y
Linearity lot verication 0: display only
1: yes/no conrmation
2: prompt for lot entry
1 Y Y
Display
Contrast 1 - 15 7 Y Y
Formats and Language
Date format 1: MM/DD/YY
2: DD.MM.YY
1 Y Y
Time format 1: 24 hours
2: 12 hours
2 Y Y
Language setting * 1: German
3: French
4: Spanish
5: Italian
6: Dutch
7: Swedish
8: English (USA)
9: Danish
11: Portuguese
8 Y Y
Glucose result units 0: mg/dL
1: mmol/L
0 Y Y
* Other languages: For the availability of languages not listed above, contact Roche.
Power Up
Location (a string to show where a certain
meter ought to be located); to be shown on
the Power Up screen.
0 – 20 characters “” N Y
Meter Functionality
Date and time editing allowed 0: electronically only (only
the DMS can set date and
time)
1: anyone (anyone can set
date and time)
2: password required (the
setup password is required
to set date and time)
1 N Y
Main menu 2 “Linearity” (if enabled,
linearity tests are allowed)
0: disabled
1: enable
1 Y Y
Main menu 2 “Maintenance” (if enabled,
maintenance comments are allowed)
0: disabled
1: enable
1 Y Y
Zobrazit stránku 146
1 2 ... 142 143 144 145 146 147 148 149 150 151 152 ... 191 192

Komentáře k této Příručce