Roche Accu-Chek Aviva Technické informace Strana 238

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 282
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 237
236
Display Shows Possible Cause(s) Possible Solution(s)
E-53 Bad Test
Error
j A meter or test strip error has
occurred.
j Discard the test strip and repeat the test.
j Your blood glucose may be
extremely high.
j If this matches how you feel, contact your
healthcare professional immediately. If it does not
match how you feel, repeat the blood glucose test
and see Chapter 2, Testing Your Blood Glucose.” If
this still does not match the way you feel, run a
control test with your control solution and a new
test strip. If the control result is within the
acceptable range, review the proper testing
procedure and repeat your blood glucose test with a
new test strip. If the E-53 error still appears for your
blood glucose test, your blood glucose test result
may be extremely high and above the system’s
reading range. Contact your healthcare
professional immediately. If the control result is
not within the acceptable range, see Chapter 3,
“Control Testing.”
50813_00048009694_D.indb 236 7/3/12 3:13 PM
Zobrazit stránku 237
1 2 ... 233 234 235 236 237 238 239 240 241 242 243 ... 281 282