Display Shows Possible Cause(s) Possible Solution(s)
Test Strips Expiring j The test strips expire at the
end of the current month.
j This message may appear when using a white
activation chip in the meter. It means the test strips
expire at the end of the current month. At the end of
the month, discard the white activation chip and
any remaining test strips. Insert a black activation
chip. Make sure the time and date in the meter are
correct. If you need a black activation chip, please
contact Roche.
j Ensure the time and date are correct and adjust, if
necessary. See Chapter 5, “Changing Meter
Settings.
Time Mismatch j The time or date on the meter
is more than 5 minutes
dierent than the time or date
on the pump.
j The meter time and date were changed to match
the pump time. Ensure the time and date are
correct. If the time or date are not correct, change
the time and date on the pump.
i NOTE
For additional information on errors and warnings, see Chapter 6, “Icons, Reminders, Warnings, and
Errors.”
• If you have a question or if you see any other error screen, please contact Roche.