Meter Setup
58
System Administrator (Admin.)
In the default setting
, the meter is not protected with an Admin. password, and all setup options
are accessible to every user. If you set up an Admin. password, the following setup areas are
automatically reserved only for the system administrator/supervisor (i.e., the person who knows
the password).
Screen: Result Units, Date/Time
Options: Connection (to a computer or a printer)
ID Setup (the entire area)
QC Lockout
QC Range
Operator Lockout (only available in combination with a data management system)
When you enter an Admin. password, this password will have to be entered from this point
forward before any of the settings above can be changed. The Admin. password must also be
entered before you can delete or change the Admin. password itself. If you forget the Admin.
password, contact your Roche representative.
Zobrazit stránku 57