Testing a Blood Sample
92
8 The test strip icon prompts y
ou to insert a
test strip. Remove a test strip from its con-
tainer and close the container again with
the stopper.
Exposure to external influences (such as
humidity) may deteriorate the test strips and
may lead to error messages. Therefore, always
close the strip container immediately after
removing a test strip.
9 Hold the test strip so the lettering with the
test name is facing upward.
10 Slide the test strip into the test strip guide
in the direction indicated by the arrows.
Slide the test strip in as far as it will go. A
beep indicates that the meter has
detected the test strip (provided the
beeper is enabled).
If you use a new test strip lot and have not
inserted the code chip yet, you must do so
now. Otherwise you cannot perform a test.
The meter displays the number of the
code chip belonging to the new test strip
lot. Depending on the setting, you may
also be required to run a liquid quality
control test at this point.
Test 09:15 am
04/19/2012
P-ID: PID01
Müller, David
Op: Schulz
Par: PT
Test 09:15 am
04/19/2012
184
P-ID: PID01
Smith, Henry
Op: Operator 1
Par: PT