ID Setup
Use the ID Setup screen to enter settings for user management and patient management.
These settings are optional; the meter can be operated without these settings.
There are three types of identification used with the CoaguChek XS Plus meter:
n The System Administrator, who has special rights to enter certain meter settings (provided
the administrator has been set up), and is the only one who can enter and change these
settings. It is not necessary for Administrator identification to be activated to use the
CoaguChek XS Plus meter. However, it can be desired or necessary, depending on the
environment and site of use.
n The Operator, who uses the meter to test patients. Operators are defined in the CoaguChek
XS Plus meter using external software (i.e., in operator lists managed by the facility’s
meter administrator. Administrators assign operator passwords and expiration dates and
may download these lists into the meter). This option can be helpful in a clinical setting
to provide further documentation and to ensure that the meter is used only by qualified
personnel.
n The Patient, whose test results are recorded. You can block input of a unique patient ID
(in this case, every test is simply numbered in consecutive order), permit it, or require it for
every test.
Meter setup
40
ID Setup
09:15 am
09/16/2009
Patient
Operator
Admin.
The screen prompts in ID Setup show what the
current settings are, and how the particular area is
set up:
n The standard display of Admin. means the
function is available but not activated (an Admin.
password has not been assigned).
n When the text in the Operator prompt is displayed
in gray, this means an operator list has not been
loaded (from the PC). The function is not available.
n When the Patient button has a black background,
this means the function is available and activated
(either as an option or a requirement).