Gather what you need:
n CoaguChek XS Plus meter
n CoaguChek XS PT Test Strips and matching code chip
n CoaguChek Lancet
n CoaguChek Capillary tube/bulb (optional)
Do not use glass capillary tubes or capillary tubes that contain anticoagulants.
n Cotton ball and alcohol wipe
Important notes
Always …
n … close the container immediately after having removed a test strip.
n … operate the meter at an ambient temperature between 59°F and 90°F (15°C and 32°C).
n … place the meter on a level, stable surface (table) or hold it so it is roughly horizontal.
n … follow the information on correct handling of test strips in the package insert.
n … keep the test strip guide and housing clean. See the chapter entitled Cleaning.
Never …
n … store the meter at extreme temperatures (greater than 104°F or 40°C).
n … store the meter in damp or humid conditions without protection.
n … remove or insert the code chip while the meter is performing a test.
n … use the code chip from a pack other than the one in use.
n … touch or remove the test strip during a test.
n … wait more than 15 seconds after lancing the fingertip before applying the blood.
n … add more blood after the test has begun.
n … perform a test with a drop of blood from a previous puncture.
Testing a capillary blood sample
Testing a capillary blood sample
53
When collecting samples always observe the general precautions and
guidelines related to blood sampling.
Dispose of all test strips used for patient testing in line with the disposal
policy of your laboratory or practice.
Failure to comply with the above may lead to inaccurate results.